![Pyramid Technical Consultants F3200E User Manual Download Page 37](http://html1.mh-extra.com/html/pyramid-technical-consultants/f3200e/f3200e_user-manual_1632602037.webp)
PSI System Controls and Diagnostics
F3200E User Manual
F3200E_UM_180726 Page 37 of 107
Figure 19. Reading background noise in internal trigger mode
11.3 Screen layout
The F3200E user interface screen is divided into two halves plus a top banner area.
The top banner contains the following indicators and controls:
Comms bar
When moving, this indicates that messages from the F3200E are being
received by the PTCDiagnosticG2. The message rate is indicated,
typically about 20 Hz.
Connected LED
When lit, this indicates that communications are valid and the system is
not in error.
Busy LED
When lit, this indicates the F3200E is busy and cannot respond to inputs,
for example while performing a calibration.
Measuring LED
When lit, this indicates that data acquisition has been initiated.
Error LED
When lit, an F3200E error has been latched. There will be a
corresponding red error message in the message window. The error can
be acknowledged and cleared with the Clear last error button below the
message area
.
Auto Initiate
If this box is checked, the F3200E will start a new acquisition whenever
any parameter is altered.