P/N: M-145MAT00-20
Rev. A -
Page Rev.A
III-21
3
-b) Calibration
1-Replace the test kit with the calibrated output (figure 3.2 page III-4)
and block
the output
2-With the mouse, click on the setting cursor of the menu “Flow/Débit” and adjust
the
Offset
to
0
,
3-Store with the button “Storage”,
4-Place TR4 in a medium position,
5-Adjust TR3 for a
Value/Valeur
reading of
0 l/s +/-0.05
(Value of Flow in the
“Flow/Débit” menu),
6-Leave the calibrated output free and adjust TR4 for a
Value/Valeur
reading of:
1.2 l/s +/-0.05
,
7-Block again the calibrated output and verify that the value reading is:
0 l/s +/-0.05.
If not, repeat the process from the step #5.
CAUTION
Allow to warm up for 1/2 hour with a set pressure of 14cmH
2
O, unit in Cpap
mode, before carrying out the pressure calibration.
Setting Cursor
{
Menu
"
Flow/Débit
"
GoodKnight 425 / 425 ST
Service Manual