P/N: M-145MAT00-20
Rev. A -
Page Rev.A
III-20
3) Flow sensor calibration
3-a) Set up material
1-Remove the top housing (refer to the disassembly section page III-22),
2-Locate on the MCU board the trimmers TR3 and TR4,
3-Realise the test set up figure 3.1 page III-4
4-Connect the unit to the computer,
5-Follow the steps described page III-5 to III-6 and click on “Calibrations/Performances”,
the following menu will appear.
TR4
TR3
GoodKnight 425 / 425 ST
Service Manual