4) Preparation for testing
a
- Verification of the pressure
b
- Verification of the flow measurement chain
P/N: M-145MAT00-20
Rev. A -
Page Rev.A
III-4
Figure 3-2
Calibrated output M-BH0210
Figure 3-1
Pressure meter
PTS 2000 P/N: 4-076185-00
Patient tube M-261000-27
Test Kit
with 4 mm leak
P/N: M-212430-02
(including coupling)
Coupling
Connecting tube
M-213530-05
GoodKnight 425 / 425 ST
Service Manual