PUREGAS, LLC
P200W2 Series Dehydrator User’s Guide
Page 40 of 82
P012213 – Rev. F
8.8.3
Press the Up (
) Arrow Button to access the next screen.
8.8.4
Set Low Pressure Alarm Point
(default setting is 1.0 KPa) –
8.8.4.1
Press the Function (
)
Button to Set.
8.8.4.2
Press the Left (
) &
Right (
) Arrow Buttons
to move the underscore
beneath the digit to change.
8.8.4.3
Press the Up (
) & Down (
) Arrow Buttons to change the value of
the selected digit.
8.8.4.4
Press the Function (
) Button when Done.
8.8.4.5
Press the Left (
) &
Right (
) Arrow Buttons
to move the underscore
beneath the correct
confirmation choice (
Y
es or
N
o).
8.8.4.6
Press the Function (
) Button to Confirm. This will lock in the new
setting value.
8.8.5
Press the Up (
) Arrow Button to access the next screen.