NORMI 13 (T42023)
Intended Use - Accessories
D703.131.00/03 en
7
Intended Use
NORMI 13 (T42023) is a test object for constancy
tests of digital projection radiography systems
according to DIN 6868-13.
The standard DIN 6868-13 applies to equipment
working with imaging plates or semiconductor
image receptors and where the diagnosis is car-
ried out using a film at the viewing box or image
display device (diagnosis on a clinical review
monitor). It describes a method that helps to test
the constancy of decisive performance character-
istics of X-ray equipment.
NORMI 13 fulfills all requirements of the standard
DIN 6868-13 for constancy tests.
This user manual describes the design of the test
object, the test setup, and the test procedure. The
evaluation of the results is the sole responsibility
of the user.
HINT
This user manual only describes the use of the
NORMI 13 test object. For detailed information
about constancy tests, please refer to the stan-
dard DIN 6868-13.
WARNING
Hazards originating from other system compo-
nents.
Hazards to Persons! Equipment Damage!
Observe the safety information provided in the
user manuals of the medical X-ray equipment
and of the measuring devices.
NORMI 13 (T42023) will be referred to as test
object in the following chapters.
Note:
In the United States, federal law restricts
this device to sale by or on the order of
a physician.
Accessories
The following accessories are available for the
test object:
– Bucky mounting device for diagnostic test
object T20005:
for the convenient fixation of NORMI 13 and a
PMMA absorber to the wall bucky
– Wall stand holder T20018:
for the convenient fixation of NORMI 13 to
the wall bucky, with or without an additional,
optional copper plate. The wall stand holder
must always be used in combination with the
bucky mounting device T20005.
For more information about using the acces-
sories, refer to section