Security Information
[D151.131.0/3]
Operation in the patient environment
•
The product represents a medical system. The
product may be operated in the patient
environment.
Exclusion of operation as device with patient
contact
•
Neither the device nor any peripheral device may
have contact to the patient.
Exclusion of operation as controlling instrument
The device is only for use to measure the dose
within radiation field. The product must not be used
to control radiodiagnostic equipments.
Safety Instruction to avoid electrical hazard
•
The product is a device of IEC 601 protection
class II.
•
Before connecting the device to the power line,
make sure that line voltage and frequency are
equivalent to the ratings indicated on the
nameplate of the device.
•
Before putting the device into operation, visually
inspect all connection cables for signs of
damage. Damaged cables and connectors must
be replaced immediately.
Never touch the patient and open connectors of the
device at the same time.
This symbol stands for:
Please observe the manual !