Connection and Care of the Ionization
Chamber
[D151.131.0/3]
5
Connection and Care of the Ionization Chamber
The connection between the ionization chamber and the DIAMENTOR E is performed by the
cable T26004. This is a special cable with cable core and three concentric shieldings. The metal
parts of plug and socket are connected to the outer shielding which has mass potential. When
installing the cable, it must be sure that plug and socket of the cable do not touch metallic parts
of the X-ray machine, to avoid current loops.
Do not clean the surfaces of the ionization chambers with alcohol or other solvent materials.
Use only a moistened cloth or a tissue with one of the following disinfectants:
1.
2%
Lysoform
2.
2%
Korsolin
3.
1%
watery solution of Sokrena
4.
5 %
Sagrotan
5
5 %
Gigasept