Frequency:
Always clean after each use. Always clean and disinfect before storing
and between each user. Please read the disinfectant manufacturer’s product label.
Recommended Disinfectant:
Super Sani-Cloth® wipes only or any disinfectant with
the same EPA Reg. No. (EPA Reg. No. 9480-4, Professional Disposables International, Inc.
(PDI), Orangeburg, NY), concentration of active ingredients (0.25%) and with a contact
time of 2 minutes. The active ingredients in this disinfectant are n-Alkyl dimethyl
ethylbenzyl ammonium chlorides. Super Sani-Cloth was tested and found to be
effective per recommended guidelines when used with this system.
Please only use
this disinfectant. Use of other disinfectants may cause damage to your
analyzer. Do not use bleach, peroxide, or window cleaners on this analyzer.
If you have any questions or need to know where to purchase the disinfectant wipes,
call PTS Diagnostics Customer Service at 1-877-870-5610 (US) or +1-317-870-5610.
There are a large number of distributors of this disinfectant. If you cannot obtain from
the distributor who supplies your other supplies, please contact us for assistance.
A
B
C
D
E
The entire case surface should be cleaned and disinfected.
Cleaning Instructions
Cleaning removes visible soil, organic material, and most importantly, blood products.
Always clean
before
disinfecting.
1.
Please see picture above. Clean and disinfect all surfaces of this analyzer.
2.
Obtain recommended wipes.
3.
Using a fresh wipe, wring out excess liquid and carefully wipe to clean.
4.
Allow to air dry or dry with cotton gauze.