Alarm type
Message
Alarm priority
Inspiratory air Flow Sensor is failure!!!
High
Inspiratory O
2
Flow Sensor is failure!!!
High
Table 2-1. Alarm information and priority
Caution:
When different priority of alarms occur simultaneously, only the alarm with highest
priority will display.
When the same priority alarms occur simultaneously, the current alarms are
displayed alternately on the alarm information area.
If high priority alarms and medium priority alarms occur simultaneously, the alarm
indicator light flashes in red.
If multiple alarms occur, the alarm messages are displayed in order of priority and
time of occurrence.
2.10.3 Alarm limit setting
Refer to
5.9 alarm setting.
Press the
【
Alarm setting
】
to enter the alarm setting screen
and set the alarm limit value.
Warning:
A HAZARD can exist if different ALARM PRESETS are used for the same or similar
equipment in any single area, e.g. an intensive care unit or cardiac operating theatre.
Caution:
Please always pay attention to the alarm limit parameter and make sure it is set at a
proper and reasonable level to avoid inaccurate alarms.
When mains power is interrupted, the alarm limit setting does not change and the
system saves the latest setting before interruption.