79
Parameter
Description
Accurancy
P
peak
(pressure support) = pressure setting
(Pressure support) + PEEP
±2cmH
2
O or ±10% of
the set value,
whichever is greater
Pressure control
Range: 5
~
70 cmH
2
O
Increment: 1 cmH
2
O
P
peak
(pressure control) = pressure setting
(pressure control) + PEEP
High pressure
level(optional)
Range: 5
~
70 cmH
2
O
Increment: 1 cmH
2
O
Not suitable for C series, and optional for D series
Low pressure
level(optional)
Range: 0
~
50 cmH
2
O
Increment: 1 cmH
2
O
Not suitable for C series, and optional for D series
High pressure
support(optional)
Range: 1
~
70 cmH
2
O
Increment: 1 cmH
2
O
Low pressure
support(optional)
Range: 1
~
70 cmH
2
O
Increment: 1 cmH
2
O
Oxygen concentration
Range: 21
~
100%
Increment: 1%
±6% of the set value
Apnea time
Range: 10
~
60 sec
Increment: 5 sec
±1 s or ±10% of the set
value
Respiration sensitivity
Range: 10
~
60%
Increment: 10%
P
rate
Range: 1
~
5
Increment: 1
B.4 Monitoring parameters
Parameter
Description
Inspiration tidal volume
Range: 0
~
3000 mL, Resolution: 1 mL
Expiration tidal volume
Range: 0
~
3000 mL, Resolution: 1 mL
Minute ventilation volume
Range: 0
~
99 L, Resolution: 0.1 L
Spontaneous minute ventilation
volume
Range: 0
~
99 L, Resolution: 0.1 L
Total respiration frequency
Range: 0
~
100bpm, Resolution: 1 bpm
Spontaneous respiration
requency
Range: 0
~
99 bpm, Resolution: 1 bpm
Inspiration and expiration rate
Range: 4:1
~
1:9, Resolution: 0.1
PEEP
Range: 0
~
50 cmH
2
O, Resolution: 1 cmH
2
O
Airway peek pressure
Range: 0
~
100 cmH
2
O, Resolution: 1 cmH
2
O
Average airway pressure
Range: 0
~
100 cmH
2
O, Resolution: 1 cmH
2
O