Warnings:
from the ventilator and the oxygen tube. Do not use the oxygen tube which is worn or
effused or contaminated by the flammable liquid (such as grease or oil). In the
environment of high oxygen concentration, textiles, oils and other combustible
material are easy to cause fire. When you detect a burning odor, cut off the oxygen
supply device, the power supply and the spare resource immediately.
The ventilator is not designed for the MR environment that is suitable for the
Medication reaction (MR) checking. Otherwise the system may lose some functions
and it may cause permanent damage to the ventilator system.
This equipment can only be used in the specified environment, and it cannot be used
near the flammable or explosive resources and cannot be used in the mobile phone,
radioactive rays or the MRI equipment environment.
If the external interface at the rear of the ventilator is connected to the other
equipment, leakage currents may increase.
Always perform a regular clean and pre-use check before opening the package.
The equipment must only be connected to a supply mains with protective earth to
avoid electric shock.
During the usage of equipment, please do not pull display screen wire, otherwise, it
may result in alarm sound failure.
When adding attachments or other components or sub-assemblies to the breathing
system, the pressure gradient across the breathing system, measured with respect to
the patient connection port, may increase.
Operator will have to ensure that the inspiratory and expiratory resistances measured
at the patient connection port during spontaneous breathing and normal operation
shall not exceed 0.6 kPa at 60 l/min for adult use, 30 l/min for pediatric use.
Please double-check the power cord before connected to the electrical outlet, make
sure it has no damage, scratch or other factors which will lead to the inside conductor
exposed issues.
When there is any doubt about the integrality of the outside grounded protection or the