10
OPTIMAX
®
2010
Installation
Installation
1. Requirements for installation
a.
Fresh water connection: Shut-off tap, thread with 3/4" outside diameter (wash-
ing machine connection), Water pressure 2 - 10 bar (29 - 145 psi).
b.
Drainage connection: Plastic tube - inner diameter 50 mm (2”) or larger. A
ventilated syphon which serves as odour preventor should be included in the
planning. The drainage tubes should be installed with a decline of minimum
5 %.
Local Water Authorities regulations should be complied with.
c.
Electrical connection: Fused wall socket with earth connection according to
electrical data (see technical specifications, page 9). It is also required to
install an earth-leakage switch (with 25 A / 30 mA nominal error current).
2. Transport
Due to the weight and dimensions of the film processor, OPTIMAX
®
2010 should
always be carried by two persons. To do so, hold the machine at the sides on the
bottom (see figure). While putting the machine down, watch the position of the lev-
elling feet to prevent damaging these.
3. Set-Up
Unpack the Processor. Remove cover and transport securing brackets on the sides
of the roller racks. Remove roller racks - start with the dryer rack.
The standard version of OPTIMAX
®
2010 is delivered as a tabletop processor with
a three part base plate. If the machine is upgraded with the optional stand or cabi-
net, the two side parts must be removed from the base plate.
Table-top installation
If the unit is to be set up on a table top or work table, only the four unit feet need to
be screwed out and levelled..
Installing on processor stand or base cabinet
In the event that the processor is to be installed on the stand or cabinet (optional
accessory), the processor will be mounted directly to it. Mount processor according
to manual included with stand or cabinet (the adjustable feet inside the accessory
bag are not required). Remove the two side parts of the base plate before.
Finally the processor needs to be levelled:
Place spirit level across the sidewalls of processor and adjust the levelling feet ac-
cordingly. Replace the racks into the processor and close the latches.
Electrical connections should be carried out according to reg-
ulations by an electrician.
Attention!
Machine should not be installed on table-top without adjusta-
ble feet. as this would block the ventilation openings under the
machine and cause overheating.