8
OPTIMAX
®
2010
Technical Specifications
Intended Use
The X-ray film processor OPTIMAX
®
2010 is intended exclusively for the purpose
describe in the introduction above.
X-ray film processors (MDD class I) are employed in “medical” applications (medi-
cal products directive) and “non-medical” applications (electrical appliances and
EMC directives).
Intended use includes observing the operating instructions, the installation instruc-
tions and adherence to the safety notes.
Any application differing from intended use voids the guarantee by PROTEC
®
.
The owner of the machine will be liable for damages resulting from unintended use
or faulty application.
Intended use includes adherence to all statutory regulations concerning occupa-
tional safety and radiation protection applicable at the operating site.
Technical Specifications
Film transport:
Continuous roller transport system
Film formats:
In general: Sheet and roll films up to 35.8 cm
(14.1 ’’) width;
Roll films with leader from 70 mm (2.8 ’’) width;
Smallest film format 10x10 cm (4x4 ’’).
Mammography type 1161: For processing mam-
mography films.
Graphics-art type 1162: With cassette box
(LxWxH) 35x13x12 cm (13.8x5.1x4.7’’) for
processing roll films.
Processing capacity:
129 films 24x30 cm (10x12 “) per hour
(standard unit, film transverse, at 90 s)
Process time:
1.5-3 min.; adjustable in steps of 0.1-min.
Linear speed:
28-56 cm/min., depending on selected cycle time.
Developing time:
25-49 s, depending on selected cycle time.
Tank capacities:
Developer, fixer and water tank 5 litres each
Circulation system:
Developer and fixer are continuously circulated by
a circulation pump
Replenishment:
Automatic by film surface measurement in relation
to processed film; replenishment can be switched
off; time replenishment can be activated.
Developer temperature:
Adjustable 28 - 37 °C (82.4 - 98.6 °F)
Fixer temperature:
Adjusted to developer temperature by heat
exchanger.
Dryer temperature:
Adjustable between 10 - 99% of dryer output, tem-
perature achieved depends on line voltage.
Water connection:
Permissible water pressure 2 - 10 bar (29 - 145
psi), permissible water temperature 5 - 30 °C (41 -
86 °F).
Water consumption:
1.9 litre per minute when processing.
Water saving mode: 0.9 litres per minute
Drain capacity:
7 litres per minute
Noise level:
Less than 58 dB(A).
Heat emission:
During processing approx. 1.4 kJ/s.