V6.0_2021-05-25
OPTIMAX 2010 Instructions for use
5160-0-0002
PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
31 of 35
7.1
Dimensions
7.2
Protection Type and Protection Class
The X-ray film processor complies with protection class I.
7.3
Environmental Conditions
7.3.1
Environmental Conditions during Operation
Usage
Only indoor (darkroom)
Surrounding temperature
18 - 40 °C (51,6 - 104 °F), ventilated room, room
temperature must be lower than set bath temperature.
If the ambient temperature exceeds the set bath
temperature, the image quality may deteriorate.
Relative humidity
Humidity less than 80% up to 31°C (80°F), decreasing
linearly to 50% at 40°C (104°F).
Air pressure
Altitude above sea level less than 2000m (6666 feet).
7.3.2
Environmental Conditions during Shipping and Storage
Surrounding temperature
- 10°C to +70°C
Relative humidity
10% to 95% (non-condensing)
Air pressure
500 hPa to 1060 hPa