29
MARK76X5002 Rev 3
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Luxura X5 Model 38X
is intended for use in the electromagnetic environment specified below.
The customer or the user of the Luxura X5 Model 38X should assure that it is used in such an
environment.
Immunity Test
IEC 60601 Test Level
Compliance Test
Electromagnetic
Environment - Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±2 kV, ±4 kV & ± 8 kV for
Contact Discharge
±2 kV, ±4 kV, ±8 kV and±
15kV for Air Discharge
±2 kV, ±4 kV & ± 8 kV for
Contact Discharge
±2 kV, ±4 kV, ±8 kV and±
15kV for Air Discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
±2kV for power supply lines
±1kV for input/output lines
±2kV for power supply lines
±1kV for input/output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5
AC Mains Line to Ground
±0.5kV, ±1kV and ±2kV
AC Mains Line to Line
±0.5kV and ±1kV
AC Mains Line to Ground
±0.5kV, ±1kV and ±2kV
AC Mains Line to Line
±0.5kV and ±1kV
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% UT (100% dip in UT)
for 0.5 cycles
0% UT (100% dip in UT) for
1 cycle
70% UT (30% dip in UT) for
25/30 cycles
<5% UT (100% dip in UT)
for 250/300 cycles
<5% UT (100% dip in UT)
for 0.5 cycles
0% UT (100% dip in UT) for
1 cycle
70% UT (30% dip in UT) for
25/30 cycles
<5% UT (100% dip in UT)
for 250/300 cycles
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Luxura X5 model
38X requires continued
operation during power
mains interruptions, it
is recommended that
the Luxura X5 model
38X be powered from an
uninterrupted power supply
or a battery.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE: UT is the a.c mains voltage prior to application of the test level.
Summary of Contents for Luxura X5 38 SLi High Intensive
Page 1: ...Luxura X5 38 SLi High Intensive 230V User Manual Model 38X 12 MINUTES MARK76X5002 Rev 3...
Page 66: ...66 MARK76X5002 Rev 3 1 Assembly...
Page 67: ...67 MARK76X5002 Rev 3 2 3...
Page 68: ...68 MARK76X5002 Rev 3 3 4...
Page 69: ...69 MARK76X5002 Rev 3 6...
Page 70: ...70 MARK76X5002 Rev 3 7...
Page 71: ...71 MARK76X5002 Rev 3 8...
Page 72: ...72 MARK76X5002 Rev 3 9...
Page 73: ...73 MARK76X5002 Rev 3 10...
Page 74: ...74 MARK76X5002 Rev 3 11...
Page 75: ...75 MARK76X5002 Rev 3 12...
Page 76: ...76 MARK76X5002 Rev 3 13...
Page 77: ...77 MARK76X5002 Rev 3 14...
Page 78: ...78 MARK76X5002 Rev 3 15...
Page 79: ...79 MARK76X5002 Rev 3 16...
Page 80: ...80 MARK76X5002 Rev 3 17 18...
Page 81: ...81 MARK76X5002 Rev 3 19...
Page 82: ...82 MARK76X5002 Rev 3 20...
Page 83: ...83 MARK76X5002 Rev 3 21...
Page 84: ...84 MARK76X5002 Rev 3 22 23...
Page 85: ...85 MARK76X5002 Rev 3 24...
Page 86: ...86 MARK76X5002 Rev 3 25 26...
Page 87: ...87 MARK76X5002 Rev 3 27 28...
Page 88: ...88 MARK76X5002 Rev 3 29...
Page 89: ...89 MARK76X5002 Rev 3 30...
Page 90: ...90 MARK76X5002 Rev 3 32 31...
Page 91: ...91 MARK76X5002 Rev 3 33 34...
Page 92: ...92 MARK76X5002 Rev 3 35...
Page 93: ...93 MARK76X5002 Rev 3 36...
Page 94: ...94 MARK76X5002 Rev 3 37 38...
Page 95: ...95 MARK76X5002 Rev 3 39 40...
Page 96: ...96 MARK76X5002 Rev 3 41 42...
Page 97: ...97 MARK76X5002 Rev 3 43...
Page 98: ...98 MARK76X5002 Rev 3 44 45...
Page 99: ...99 MARK76X5002 Rev 3 46 47...