MARK76X5002 Rev 3
Technical Data
Electrical Specifications
Luxura X5 38 SLI High Intensive
Power consumption rating:
32 Amp
Electrical Rating:
230V 3~ 40A 60Hz
Rated Fuses:
40 amp 3-pole
Continuous Operation:
Continuous operation for a maximum of
12
minutes.
Classification:
Type B Applied Parts
Ingress protection classification:
IP20
Protection against electric shock:
Class I ME Equipment
Noise level inside the device:
71 DBC
Noise level 1m away from the device:
62 DBC
Load on the base plate:
Max 250lbs or 113 Kg
Equipment Weight:
Approx 845 lb or 383 kg
Operating Humidity:
30-75%
Ambient Operating temperature:
50 to 104 degrees Farenheit
10 to 40 degress Celcius
Max. operating altitude (above sea level): <6500 ft (2000m)
Max surface temperature
146.6 degrees Farenheit
63.7 degrees Celcius
Note: If operating this equipment
at altitudes above 6500 ft / 2000m
above sea level adaption may be
needed to get normal operation.
Summary of Contents for Luxura X5 38 SLi High Intensive
Page 1: ...Luxura X5 38 SLi High Intensive 230V User Manual Model 38X 12 MINUTES MARK76X5002 Rev 3...
Page 66: ...66 MARK76X5002 Rev 3 1 Assembly...
Page 67: ...67 MARK76X5002 Rev 3 2 3...
Page 68: ...68 MARK76X5002 Rev 3 3 4...
Page 69: ...69 MARK76X5002 Rev 3 6...
Page 70: ...70 MARK76X5002 Rev 3 7...
Page 71: ...71 MARK76X5002 Rev 3 8...
Page 72: ...72 MARK76X5002 Rev 3 9...
Page 73: ...73 MARK76X5002 Rev 3 10...
Page 74: ...74 MARK76X5002 Rev 3 11...
Page 75: ...75 MARK76X5002 Rev 3 12...
Page 76: ...76 MARK76X5002 Rev 3 13...
Page 77: ...77 MARK76X5002 Rev 3 14...
Page 78: ...78 MARK76X5002 Rev 3 15...
Page 79: ...79 MARK76X5002 Rev 3 16...
Page 80: ...80 MARK76X5002 Rev 3 17 18...
Page 81: ...81 MARK76X5002 Rev 3 19...
Page 82: ...82 MARK76X5002 Rev 3 20...
Page 83: ...83 MARK76X5002 Rev 3 21...
Page 84: ...84 MARK76X5002 Rev 3 22 23...
Page 85: ...85 MARK76X5002 Rev 3 24...
Page 86: ...86 MARK76X5002 Rev 3 25 26...
Page 87: ...87 MARK76X5002 Rev 3 27 28...
Page 88: ...88 MARK76X5002 Rev 3 29...
Page 89: ...89 MARK76X5002 Rev 3 30...
Page 90: ...90 MARK76X5002 Rev 3 32 31...
Page 91: ...91 MARK76X5002 Rev 3 33 34...
Page 92: ...92 MARK76X5002 Rev 3 35...
Page 93: ...93 MARK76X5002 Rev 3 36...
Page 94: ...94 MARK76X5002 Rev 3 37 38...
Page 95: ...95 MARK76X5002 Rev 3 39 40...
Page 96: ...96 MARK76X5002 Rev 3 41 42...
Page 97: ...97 MARK76X5002 Rev 3 43...
Page 98: ...98 MARK76X5002 Rev 3 44 45...
Page 99: ...99 MARK76X5002 Rev 3 46 47...