Technical data
Measured variables
Peracetic acid
Applications
Particularly suitable for Cleaning in Place (CIP), rinser, PET bottle
cleaning
Measuring range
PAA 1-mA-200 ppm:
1… 200 mg/l
PAA 1-mA-2000 ppm: 10… 2000 mg/l
Dispersal
In accordance with the lower limit of the instrument measuring
range
Nominal slope conductance
With pH 4, T = 30
°
C:
PAA 1-mA-200 ppm:
60
µ
A/ppm
PAA 1-mA-2000 ppm: 6
µ
A/ppm
Reaction time
T
90
approx. 3 min
pH range
1 - 9 (peracetic acid stability range)
Temperature range
5 - 45
°
C, temperature compensated, no sudden leaps in
temperature (sample water), 5 - 50
°
C (air temperature)
Pressure
Sample water in DLG III in-line probe housing:
Maximum 1 bar (free flow)
Sample water in DGM in-line probe housing:
Maximum 3 bar (30
o
C, no sharp pressure peaks permitted)
Flow
Sample water through in-line probe housing DLG III, DGM
optimal:
50 l/h minimum: 20 l/h
maximum: 100 l/h
Cross-sensitivity
Bromine, bromamine, chloride, ozone.
The probe shows no signs of cross-sensitivity to standard aqueous
solutions of hydrogen peroxide.
Lifespan of diaphragm cap
Typically 3-6 months, depending on water quality
Materials
Diaphragm cap:
PVDF
Electrode shaft:
PVC-C
Power supply
16 - 24 V DC; min 35 mA at 16 V DC
Output signal
4 - 20 mA
Enclosure rating
IP 65
Storage temperature
Between 5 and 50
°
C
Disposal / Technical Data