6) coating/deposits (mangan-
Clean or change diaphragm
ese, iron oxide) on the
cap (see sect. 6);
diaphragm cap
run in the measuring cell (see
sect. 8.1),
calibrate (see sect. 8.2)
Measuring cell measured
No electrolyte in the
Fill with electrolyte (see sect. 6
value is 0 ppm
diaphragm cap
assembly, sect. 8.1 running-in
time and sect. 8.2 calibration)
Measuring cell measured
1) measuring cell connected
Connect the measuring cell
value is 0 ppm and error
to controller with incorrect
correctly to the controller
message appears on the
polarity
(see sect. 7)
DULCOMETER
®
D1C
2) signal cable broken
Change signal cable
controller “PES input < 3 mA”
3) defective measuring cell
Return the measuring cell
4) defective control equipment
Check the control equipment
with measuring cell simulator
(DULCOMETER
®
Simulator,
order no.1004042),
return if faulty
Measuring cell measured
1) running-in time too short
See section 8.1 running-in
value is 0 ppm and measuring
time
cell current is between 3.0
2) defective reference
Return the measuring cell
and 4.0 mA**
electrode*
for regeneration
Error message on
1) PES content exceeds
Check system, fix the fault,
DULCOMETER
®
D1C
upper limit of instrument
repeat calibration (see sect.
controller “PES input >23 mA“
measuring range
8.2)
2) defective measuring cell
Return the measuring cell
Measuring cell measured
1) defective reference
Return the measuring cell for
value is unstable
electrode*
regeneration
2) restricted operation
Optimise control operation
* If the reference electrode has a silvery sheen or looks white, it needs to be regenerated.
Brownish-grey discolouration is however normal.
** To isolate faults, the current in the measuring cell can be displayed via the DULCOMETER
®
D1C whilst the latter is electrically connected to the measuring cell.
You can read about this in the complete operation menu, see operating instructions manual for
DULCOMETER
®
D1C section 8 in the set-up menu “PES calibration” the value under “zero
point”. You should then confirm, not with the Enter key, but leave the menu using the Back key.
Troubleshooting
Fault
Possible cause
Solution
BA_DT_097_10_08_4spr.p65
09.10.2008, 13:19 Uhr
43
Summary of Contents for DULCOTEST PAA 1-mA-200 ppm
Page 26: ...26 ProMinent...
Page 50: ...50 ProMinent...
Page 74: ...74 ProMinent...
Page 98: ...98 ProMinent...
Page 99: ...99 ProMinent...