Pr
o
Minent
®
32
14
Technical Data
Measured
variable
Free chlorine and organically bound chlorine
Area of
application
Drinking water, swimming pool water and water of a similar quality
Measuring range
CGE 2-mA-2 ppm
0.02-2.0 mg/l
CGE 2-mA-10 ppm 0.1-10 mg/l
CGE 2-4P-10 ppm
0.1-10 mg/l
pH range
5.5 - 9.5
Temperature
range
5 - 45
°
C,
Temperature compensation
No sharp rises in temperature
Storage
temperature
Frost-free between 5 and 50
°
C
Resolution
Corresponds with the value at the lower end of the measuring range
Max. pressure
DLG III A/B:
1 bar (free flow)
DGM:
3 bar (no pressure surges!)
Flow
DLG III and DGM in-line probe housing
Optimum:
30 l/h
Minimum:
20 l/h
Maximum:
100 l/h
Cross
sensitivity
Bromine, iodine, ozone, ClO
2
, chloramines and other oxidising agents lead to
measurement errors
Conductivity
Test water
30
µ
S/cm - 10.000
µ
S/cm
Operating life
Membrane cap
Typically 1 year depending upon the water quality
Materials
Membrane cap:
PPE
Electrode holder: PMMA
Electrode adaptor:
PVC
Electronics: Electronic components
Power supply
16 - 24 V DC; min 35 mA at 16 V DC
Output signal
4 - 20 mA (mA-Variant)
0 - 2 V
(4P-Variant)
Enclosure rating
IP 65
Technical Data
BA_DT_085_09_07_4spr.p65
13.09.2007, 14:18 Uhr
32
Summary of Contents for DULCOTEST CGE 2-4P-10 ppm
Page 18: ...ProMinent 18 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 18 ...
Page 34: ...ProMinent 34 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 34 ...
Page 50: ...ProMinent 50 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 50 ...
Page 66: ...ProMinent 66 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 66 ...
Page 67: ...ProMinent 67 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 67 ...