Transport and storage / Assembly
4
Transport and storage
NOTE
The probe should be transported only in its original
packaging.
Keep the packaging, including the polystyrene, and use it to
return the probe if repair becomes necessary.
Contents
•
CGE measuring cell complete with diaphragm cap.
•
Bottle containing electrolyte (50 ml)
•
Nozzle for bottle
•
Membrane cap, spare membrane cap
•
Operating instructions
•
Small screwdriver
Storage
•
Storage period for the probe including the
membrane in the original packaging
1 year
•
Storage period for the electrolyte in
the original bottle (order no. 792892)
max. 1 year
•
Storage and transport temperature
+5 to +50
°
C
•
Humidity
max. 90 % rel. humidity,
non-condensing
5
Assembly
CAUTION
Wear safety goggles and protective clothing when handling
water or solutions that contain chlorine.
IMPORTANT
•
Do not touch or damage the membrane or electrodes.
•
The electrolyte is susceptible to oxidation: always keep
the electrolyte bottle sealed after use. Do not put the
electrolyte into other transparent containers.
•
The electrolyte should not be stored for more than 1 year
and should not be yellow (see the label for the date of
manufacture).
•
Avoid air bubbles when pouring the electrolyte.
•
The membrane cap must be used only once.
BA_DT_085_09_07_4spr.p65
13.09.2007, 14:18 Uhr
24
Summary of Contents for DULCOTEST CGE 2-4P-10 ppm
Page 18: ...ProMinent 18 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 18 ...
Page 34: ...ProMinent 34 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 34 ...
Page 50: ...ProMinent 50 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 50 ...
Page 66: ...ProMinent 66 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 66 ...
Page 67: ...ProMinent 67 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 67 ...