Troubleshooting / Maintenance
When you have tried everything: check to ensure that the ring-shaped
counter electrode at the tip of the electrode stem is not a brown/grey or
yellow/greenish colour but rather silvery white. If it is the latter, it is worn out
and can be replaced by ProMinent.
9
Maintenance
IMPORTANT
•
Service the sensor regularly to avoid overdosing if it stops
working.
•
Please comply with current national regulations for
maintenance periods.
•
Do not touch the electrodes or allow them to come into
contact with greasy substances.
Service intervals
Daily/weekly depending on application.
Maintenance work
왘
Check the sensor’s reading on the controller using an appropriate
instrument for measuring chlorine (e.g. DPD-1).
왘
If necessary, recalibrate the sensor (see chapter 7.2 ”Calibration”).
Cleaning the membrane
If the membrane is contaminated and the sensor cannot be calibrated, you
can try to clean the membrane carefully.
First dismantle the sensor. Follow the safety advice.
Remove the dirt gently:
왘
Rinse the membrane under a soft, cold jet of water.
IMPORTANT
Do not use cloths, abrasive paper, scouring agents, chemical
cleaning agents or acids!
Changing the membrane
If it is not possible to calibrate the sensor even after the membrane has been
cleaned, or if the membrane is damaged, the membrane cap must be
changed (see chapter 5 ”Assembly”, chapter 7.1 ”Run-in period” and chapter
7.2 ”Calibration”).
BA_DT_085_09_07_4spr.p65
13.09.2007, 14:18 Uhr
29
Summary of Contents for DULCOTEST CGE 2-4P-10 ppm
Page 18: ...ProMinent 18 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 18 ...
Page 34: ...ProMinent 34 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 34 ...
Page 50: ...ProMinent 50 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 50 ...
Page 66: ...ProMinent 66 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 66 ...
Page 67: ...ProMinent 67 BA_DT_085_09_07_4spr p65 13 09 2007 14 18 Uhr 67 ...