DMT-pH
MeasStatus
Pos
Name
Status 0
Status 1
2
0
Measured value
Valid
Invalid
2
1
Temperature cor‐
rection
ON
OFF
2
2
Overshoot
No limit value overshoot
[MEAS_VALUE>MEAS_MA
X]
2
3
Undershoot
No limit value under‐
shoot
[MEAS_VALUE>MEAS_MA
X]
2
4
Sensor defective
OK
Sensor resistance < 2 MΩ
2
5
Input open
No error
Sensor missing
2
6
Signal input defec‐
tive
No error
Error
2
7
Sensor temp.
range
Inside
Outside (error)
Sensor type
Name
Value
with sensor monitoring
0
without sensor monitoring
1
Sensor quality
Pos
Name
Status 0
Status 1
2
0
Slope lower limit
> 45 mV/pH (OK)
40 ÷ 45 mV/pH
2
1
Slope upper limit
< 63 mV/pH (OK)
63 ÷ 65 mV/pH
2
2
Zero point lower
limit
> -50 mV (OK)
-60 ÷ -50 mV