6. Operation
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 45
6.3.3.17 Command “minimum”
command:
minimum
without
parameter
switches to minimal measurement
see also section 6.3.1.12 Key “min max” – Minimal Measurement, Maximal
Measurement
example:
minimum
display is min
6.3.3.18 Command “range”
command:
range
without
parameter
shows the current measuring range determined by probe
and gain setting
example:
range
range is 2000.0 mT
6.3.3.19 Command “relative”
command:
relative [{reference|set}]
without
parameter
show the current measuring mode and if in relative
measurement the reference value
with parameter
“set” as parameter switches to relative measurement
and takes the current measured value as reference
value.
A
number
as
parameter
switches
to
relative
measurement and uses the given number as reference
(interpreted in the currently set unit and rounded to the
current resolution).