7. Technical Specifications
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 103
7.2.13 Transverse Probe Flex 2000 mT (AS-NTP-Flex)
Figure 82 Size transverse probe Flex 2000 mT (AS-NTP-Flex)
Ranges (with FM 302)
±20 mT; ±200 mT; ±2000 mT
Effective area
2 mm
2
Thickness of support
0.6 ±0.1 mm
Length of support
70-80 mm
Width of support
4 ±0.5 mm at the tip
7 ±0.5 mm at the grip
Transfer factor
1 V/T
Bandwidth (-3 dB)
0 - 0.5 kHz
Rise time
<0.2 ms
Linearity error
<0.5 % ±0.2 mT
(0 to ±1.5 T, at 20 °C ±1 °C)
Temperature coefficient
max. -0.05 %/K, typ. -0.03 %/K (0 to 50 °C)
Zero drift
max. ±0.020 mT/K, typ. ±0.010 mT/K (DC)
Noise
typ. 7,5 µT
RMS
(10 Hz – 1 kHz)
typ. 13 µT
PP
(DC – 10 Hz, 50 s)
Operation temperature
+5 °C to +100 °C (only at first 70 mm)
+5 °C to +50 °C (grip, cable, probe connector)
Storage temperature
-10 °C to +60 °C
max. relative humidity
70 % at +35 °C
Power
±3 V through FM 302, AS-probe adapter,
AS-Adapter 3 or PLC
Connector
15 pol. SubD
Output impedance
<1
Ω
Length of cable
1.5 m
Technical specifications are subject to change without prior notice!