
7. Technical Specifications
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Page 109
7.2.19 Lot Axial Probe 200 µT (AS-UAP Lot)
Figure 88 Size Lot axial probe 200 µT (AS-UAP Lot)
Ranges (with FM 302)
±2 µT; ±20 µT; ±200 µT
Effective volume
Ø 6 mm x 25 mm
Case diameter
Ø 18.8 mm
Case length
150 mm
Transfer factor
1 V / 100 µT
Bandwidth (-3 dB)
0 – 800 Hz (2 µT- / 20 µT-range)
0 – 500 Hz (200 µT-range)
Rise time
<0.3 ms
Linearity error
<0.8 % ±0.2 µT (at 25 °C)
Temperature coefficient
max. ±0.1 %/K (10 °C to 50 °C)
Zero drift
max. ±10 nT/K
Hysteresis
max. 0.1 % of value
magnetic flux density
max. ±200 µT or 140 µT
eff
Noise
typ. 4,5 nT
RMS
(10 Hz – 1 kHz)
typ. 6 nT
PP
(DC – 10 Hz, 50 s)
Operation temperature
+5 °C to +50 °C
Storage temperature
-10 °C to +60 °C
Max. relative humidity
70 % at +35 °C
Power
±3 V through FM 302, AS-probe adapter,
AS-Adapter 3 or PLC
Connector
15 pol. SubD
Output impedance
<1
Ω
Length of cable
1.5 m
Technical specifications are subject to change without prior notice!