Page 18
7
Calibrate
7.1 Front plate with calibration standard
The front plate with calibration standard should always be mounted on the ZRS
6060. For measurements, it must be mounted in the measuring position, for
calibration in the calibration position.
Mounted in the
Removing calibration standard, Mounted in the
Measuring position turning and remounting
calibration position
If the front plate with calibration standard is
properly mounted in the measuring position, no
status message will be displayed.
If the front plate with calibration standard is not
mounted,
the status message „Plate is missing“ is
displayed. Should you carry out a measurement
without front plate, the illumination angle 0° will be
stored at the measurement.
If the front plate with calibration standard is
properly mounted in the calibration position, the
status message „Standard is mounted“ is
displayed. Should you carry out measurements
with mounted calibration standard, this
information will be stored with the measurements.
Depending on version, there are two front plates with calibration standard
– with
illumination angle (
β) +5°and with illumination angle (β) -4°
Backside of a front plate with calibration standard as an example