background image

14

 

© 2012 by Proceq SA

4.9.2 

Measuring with the Rod Electrode

The cursor begins in the top left hand corner of the XY-Grid (Fig 4.1). This indicates where the first 
measurement will be stored, so it is important to take the first measurement in the corresponding 
point marked on the concrete.

i

Note:

 The first measurement must not necessarily be made in the top left hand 

corner. Before making any measurement, you may move the cursor around on the 
screen using the 

↑↓← → 

buttons to the desired location. This is particularly useful 

if there is some kind of obstruction on the test object which means that a measure-
ment cannot be made in that position. By moving the cursor as described you can 
move on the grid to clear the obstruction and continue measuring. The important 
thing is that your position on the screen should match the actual position on the 
concrete.

• 

Moisten the foam rubber plug of the electrode with water and press it lightly onto the first mea-
suring point. The measured value will be shown in mV in the centre of the display. Once it stabi-
lizes, a beep will indicate when the measurement has been acquired automatically. (See 5.4). At 
this point the mV value is blended out and a grey scale indication will be placed on the grid. The 
cursor will move to the next point to be measured.

• 

Following the measurement it should be possible to see a wet patch when applied on dry con-
crete. If this is not the case then the foam rubber plug should be moistened with water again.

• 

The simplest way to proceed is to make the measurements as proposed by the instrument. i.e. 
begin in the top left hand corner and move along the row in the X-direction according to the 
column width, e.g. 150mm between measuring points.

• 

When you come to the end of a row, you must press the 

 button to indicate this and the cursor 

will jump down to the next row. The direction arrow (Field 2 in fig 4.1) will automatically change 
direction to 

• 

Now you can proceed to measure in the opposite direction along this row. When the end is 
reached, i.e. you are back at the Y-axis, the cursor will automatically jump down to the next row 
and the direction arrow will change back to 

.

• 

When you come to the end of a page the cursor automatically jumps to the next page.

i

Note:

 The instrument can also be used to measure along the Y-axis, i.e. up and down 

the columns. This is done by altering the direction arrow to 

 before starting. In this 

case you must indicate to the instrument that you have reached the end of a column 
by pressing the 

→ 

button.

i

Note:

 The PRINT button can be used to mark an “X” on the display in the place of a 

measurement value. This is useful for marking the position of cracks etc.

When you have finished making all of the measurements you require on the concrete press the END 
button. The measured values are stored automatically.

Summary of Contents for canin+

Page 1: ...canin more than 50 years of know how you can measure Made in Switzerland Operating Instructions Corrosion Analyzing Instrument...

Page 2: ...robe to the indicating device 8 3 3 Power ON and select the correct operating mode 9 3 4 Functional check of the equipment 10 4 Real Tests Potential Measurements 10 4 1 Planning and preparation 11 4 2...

Page 3: ...uency 27 6 12 Chipping graph 28 6 13 Annotations 28 7 Measuring the electrical resistivity 29 7 1 Preparing the concrete surface for measurement 29 7 2 Reading the display screen 30 7 3 Measuring with...

Page 4: ...correct Proceq SA makes no warranties and excludes all liability as to the completeness and or accuracy of the information 1 3 Safety Instructions The equipment is not allowed to be operated by childr...

Page 5: ...they form hydroxide OH with the generally available water and oxygen This principle creates a potential difference that can be measured by the half cell method The basic idea of the potential field m...

Page 6: ...ments Electrical resistivity of the concrete cover Measure with the Wenner Probe This topic is dealt with in detail in section 2 4 Oxygen content at the reinforcement With decreasing oxygen concentrat...

Page 7: ...rate of cor rosion decreases 2 5 Influence of Rebars on Electrical Resisitivity Measurements The presence of rebars disturbs electrical resisitivity measurements as they conduct current much better th...

Page 8: ...lphate with 100 units by weight of distilled water In order to ensure that the solution remains saturated add an additional teaspoonful of copper sulphate crystals into the electrode The electrode sho...

Page 9: ...For potential measurements the instrument must be in Corrosion analys ing mode Switch between modes by pressing MENU place the cursor at Wenner Probe press START select OFF for Corrosion Analysing mo...

Page 10: ...ect measurement on the concrete surface possible Yes No Can the influence from surface coatings be estimated and compensated Yes No Remove the surface coating 4 8 4 9 Pre moisten the surface if requir...

Page 11: ...y be used For more accessible horizontal vertical or overhead surfaces it is much faster to use the 1 Wheel Electrode For large horizontal surfaces the 4 Wheel Electrode is recommended for the reason...

Page 12: ...often used for example on the walls and ceiling of underground car parks however this can cause a small shift of the potentials It is always necessary to check whether or not a measurement can be mad...

Page 13: ...ry The number of pages that are still available for use is shown in the top right hand corner of the measuring screen Fig 5 1 e g 491 P indicates 491 pages are still available The number of pages cont...

Page 14: ...the measurement it should be possible to see a wet patch when applied on dry con crete If this is not the case then the foam rubber plug should be moistened with water again The simplest way to procee...

Page 15: ...ista i Note The nozzle 1 mm is fit as standard On dry surfaces the felt hoop should leave a clearly visible trace as you measure Depending on the character of the concrete it may be necessary to incre...

Page 16: ...Example of a typical distribution If the surface under test has both actively corroding as well as passive rebars then the two states exhibit differing statistical distributions of the potential In t...

Page 17: ...using a small grid is recommended to deter mine the actual Hotspot local potential minimum 5 General Settings 5 1 Backlight The display has a back light which can be switched on or off by pressing the...

Page 18: ...U END MENU START MENU END OFF ON END END START Main menu for resistivity measuring Pressing MENU saves the settings and takes you to the next higher menu Pressing END saves the settings and takes you...

Page 19: ...Pressing MENU saves the settings and takes you to the next higher menu Pressing END saves the settings and takes you to the measuring screen Use the and to select navigate and adjust settings Fig 5 2...

Page 20: ...alue has stabilized If automatic measuring is not selected then values have to be acquired by pressing the START button In both cases values between 200 and 50mV have to be acquired by pressing the ST...

Page 21: ...hese graphs can be inserted in an assessment report and serve the corrosion engineer together with results of non destructive and destructive tests like concrete cover depth carbonation depth chloride...

Page 22: ...Output Fig 5 2 in the Canin display device and press START Click on VistaTransfer button to start the download programme A dialog box appears see Fig 6 2 Select the respective COM port All the object...

Page 23: ...It may be that the objects from a com plete site were not all measured in the same direction See 4 9 3 ProVista allows this to be corrected Select the correct orientation and press OK again to display...

Page 24: ...ion Click on Configuration to alter the appearance and the language of the potential map Fig 6 5 Configuration Window The number range to which the colours are assigned can be changed through varying...

Page 25: ...le button and select the file you wish to add If the file is valid the next screen Fig 6 6 will allow the file to be correctly positioned The new file may be rotated or mirrored to match its orientati...

Page 26: ...ific value 6 9 Additional Functions Save File to save your changes Print View prints the currently visible section of the potential map respectively chipping graph Print File prints the entire measure...

Page 27: ...iagram 6 11 Cumulative frequency Fig 6 9 Cumulative frequency diagram Following an evaluation See 4 10 the corrosion engineer can place the 3 cursors in such a way that they represent areas of differe...

Page 28: ...ture like measured chloride content cracks or other defects can be placed directly in the potential map or the chipping graph The information will be printed and shown in bitmaps By default they are c...

Page 29: ...mmended that annotations within one file be added at different dimension levels i e zoom levels 7 Measuring the electrical resistivity Fig 7 1 Wenner four point probe with cable and control plate The...

Page 30: ...same location moving the probe a few mm between each measurement and taking a median from the 5 values The well moistened probe should be pressed lightly against the surface of the concrete until the...

Page 31: ...f the indicating device In case your PC hardware doesn t support serial ports you can use the USB converter part no 390 00 542 Preparing the PC for the Data Transfer Create directory C PROCEQ Windows...

Page 32: ...surements 980 pages 240 measurements each organised in up to 71 objects and 5 800 resistivity measurements 24 object files tables 256 measurements each Data Output RS 232 interface with USB adapter Ba...

Page 33: ...trode cable 1 5m 4 9 ft cable coil 25m 82 ft CANIN ProVista PC software on memory stick bottle with copper sul phate 250g 330 00 205 Canin Configuration with Rod and Wheel Electrodes Basic equipment s...

Page 34: ...5 Cable clip for telescopic extension 380 02 508 S Foam pad to resistance probe set of 4 pcs 11 Maintenance and Support 11 1 Functional check of the electrodes Correct function can be checked by use o...

Page 35: ...wooden plug facing upwards 11 4 Functional check of the resistivity probe See section 5 5 Assure that the Device Constant shown on the Indicating Device corresponds to the code engraved on the probe...

Page 36: ...asia proceq com Proceq Rus LLC Ul Optikov 4 korp 2 lit A Office 412 197374 St Petersburg Russia Phone Fax 7 812 448 35 00 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharjah Unite...

Reviews: