Probewell Lab Inc.
Page
•
22
Summary
The MT-1/WT3 is compatible with meter Forms 1S, 2S, 3S, 4S, 6S (36S, 46S), 8S, 9S,
12S (25S), 14S, 15S and 16S as well as
Fitzall™
meters. Before testing a meter in the
field, always make sure the electrical wiring of the meter base is fully compatible
with those shown in Appendix B.
The MT-1/WT3 socket must first be inserted between the meter base and the meter.
The installation procedure for a ringless meter base is described hereafter in Fig. 3.1.
When powered on, the MT-1/WT3 recognizes the type of meter installed and
validation of the form and the TA is required. With some meters, it is sometimes
necessary to select the form manually.
When a pickup (Optical or Metercam) is used, the socket detects the Kh of the meter
and uses it. If for some reason the Kh suggested does not match the Kh indicated on
the meter, the operator can change it manually by restarting the socket and disabling
Kh autodetection. Without a pickup, the user must manually enter the meter Kh.
The next step is to choose a type of test to perform.
If a pickup is used, the test becomes fully automatic and the test modes
available are
Custom
,
Quick,
Full, Four-Quadrant (WH) and Four-Quadrant
(VARH)*
.
Without a pickup, the modes
Manual
and
Tracking
are available. In
Tracking
,
the test is done manually and the meter disk (or the simulated disk in case of
a solid-state meter) must do a precise number of revolutions. It is up to the
operator to start and stop the test precisely pressing on START/STOP or using
the magnetic switch on the side of the socket.
Once the load is applied and the test has started, the MT-1/WT3 measures the
energy that passes through the meter with its internal electronic standard. All the
important measurements are shown on the mobile device or computer display while
testing.
When the test ends, the test result is displayed in percentage error (Ex. -0.02%) or in
percentage registration (Ex. 99.98%).
*The Four-Quadrant (WH) and the Four-Quadrant (VARH) test are only available for electronic
meters with an optical pulse.
Summary of Contents for MT-1/WT3
Page 1: ...MT 1 WT3 USER GUIDE HIGH ACCURACY THREE PHASE WIRELESS WATTHOUR METER TESTER...
Page 9: ...Probewell Lab Inc Page 8...
Page 12: ...MT 1 WT3 User Guide Page 11 Side View of the Socket Fig 2 2 Side View of the Socket...
Page 21: ...Probewell Lab Inc Page 20...
Page 43: ...Probewell Lab Inc Page 42...
Page 47: ...Probewell Lab Inc Page 46...
Page 49: ...Probewell Lab Inc Page 48 Form 1S...
Page 50: ...MT 1 WT3 User Guide Page 49 Form 2S...
Page 71: ...Probewell Lab Inc Page 70 Form 12S 25S 3N N Network...
Page 72: ...MT 1 WT3 User Guide Page 71 Form 12S 25S 3 25S Delta...
Page 73: ...Probewell Lab Inc Page 72 Form 14S 4Y s...
Page 74: ...MT 1 WT3 User Guide Page 73 Form 15S 4...
Page 75: ...Probewell Lab Inc Page 74 Form 16S 4Y...
Page 77: ...Probewell Lab Inc Page 76...
Page 83: ...Probewell Lab Inc Page 82...
Page 85: ...Probewell Lab Inc Page 84 Notes...
Page 86: ...www probewell com Probewell Lab Inc Printed in Canada Edition 2 2...