2:3 ON/OFF, PUSH BUTTONS
OK
<
>
>
To START the instrument, press OK.
The OK button is used to accept
throughout the step by step procedure.
To SWITCH OFF the instrument, press and
hold OK for 3 seconds or , when in date/time
mode, press < twice to reach the menu
alternative SWITCH OFF INSTRUMENT?
Accept by pressing OK.
Used to
change
value or
position
Used to
change
value or
position
This button
allows you to
step backwards
in the menu if
you should enter
and accept OK
by mistake.
NOTE!
If the ^ button is pushed by mistake during the measurement of a cylinder, the instrument
will step backwards in the menu with each push. It is not possible to return further back
than to CRANKSHAFT MEASURING position (see flow scheme section 5).
2:4 SETTINGS
2:4.1 Set date & time
Date and time are continuously displayed when the instrument is switched on and
controlled by an internal clock. If the battery runs empty it might be necessary to
re-enter actual time and date. Date is written YY-MM-DD (Year, Month, Date).
Time is entered using a 24 hour clock (no AM/PM). The date format can then be
changed into US format (MM-DD-YY) if desired.
2:4.2 Change number of digits
The display is normally set 3 decimal numbers (i.e. 1/1000 mm) on the display = 4
digits. It is possible to select 2 decimal numbers = 3 digits.
2:4.3 Change limit value
Red and Green indicators are located on the instruments front panel to assist
measurement. The measurement limits are normally set +/- 0.500 mm.
Measurements taken within this range will show a steady green LED.
Should it exceed these limits the Red LED will go on. If you want to make a visual
check only, the limits can easily be reduced to acceptable tolerances indicated
with Green and Red.
2:4.4 Change program mode
When selecting STANDARD MODE by OK, the program (menu) will follow the
Flow Scheme on the last page.
If MEASUREMENT ONLY is selected the display will show current date.
By pressing OK the instrument will remain in measurement only mode.
NOTE ! In this mode the panel buttons operate in the following way;
OK = ON/OFF < = RESET > = ZERO SET ^ = BACKLIGHT on/off.
This mode will be retained on re-starting the instrument and can only be switched
back to STANDARD MODE by pressing the > arrow button when the instrument is
started and displays the current date.
Step to CHANGE PROGRAM MODE and change mode.