280400 / 280401 -
A320B
- Owner’s Manual
Rev: 07-2013
Page: 6
●
Under no circumstance should the
A320B
Mobile hoist
and sling (s) be put in control of a person
who has not been properly trained in the use and care of this equipment. Failure to adhere to this warn-
ing may result in serious injury to the operator, and/or the individual being lifted/ transferred.
●
The
A320B Mobile hoist
and sling (s) are not toys. Do not use it for unsafe practices. Do
not allow children to play with the floor hoist or any of its components.
●
The manufacturer's warranty is voided if persons unauthorized by Prism Medical Ltd. will perform
work on the
A320B Mobile hoist
.
●
There are no user serviceable parts inside the actuator. Do not remove cover screws, or open the
unit, as this will
VOID THE WARRANTY.
●
In facilities where more than one operator will be responsible for using the
A320B
and sling (s),
it is imperative that all such members are to be trained in its proper use. A training program should be
established by the facility to acquaint new operators with this equipment.
●
Never expose the
A320B Mobile hoist
directly to water. Warranty does not cover any misuse or
abuse of the floor hoist system.
●
To maintain optimum function, the
A320B
should be inspected and maintained on a regular basis.
See the section titled “
General Inspection and Maintenance
”.
●
Any accessories used with the
A320B
including sling (s), should be checked to ensure that they are
in good working order. Check for signs of wear or fraying prior to use. Report any unusual wear, or
damage immediately to your local authorized Prism Medical Service Provider.
●
The
A320B Mobile hoist
and associated sling (s) are intended
only
for lifting and transferring of a
person. Prism Medical will not be responsible for any damage caused by the misuse, neglect or pur-
poseful destruction of the hoist, and/or its associated components.
●
In any circumstances do not exceed the maximum allowable load of this hoist. Refer to the
“
Specifications
” section of this manual, and/or the labels on the side of the hoist.
●
There is a risk of explosion if the hoist is used in the precsense of flammable anaesthetics.
●
Ensure that a clear space is maintained around the hoist. Move any obstacles out of the way before
opereating the floor hoist.
Cautions: