280400 / 280401 -
A320B
- Owner’s Manual
Rev: 07-2013
Page: 24
Fault Finding
Should problems arise with the use of the hoist review the following chart. Find the fault and complete the rec-
ommended solution. If the fault is not found and/or the solution does not correct the problem contact your local
Prism Medicalauthorized Service Provider for service immediately.
PROBLEM
SOLUTION
Hoist cannot raise/lower, control box is not clicking.
•
Hand control is not properly plugged in.
•
Hand control is defective.
•
Control box is defective.
Hoist cannot raise/lower, control box is clicking.
•
Lifting actuator cord (1) is not fully plugged into slot (1)
on the control box.
•
Lifting actuator is defective.
Emergency Lower button on control box does not lower.
•
Lifting actuator cord (1) is not fully plugged into slot (1)
on the control box.
•
Lifting actuator is defective.
•
Control box is defective.
Hoist cannot spread or close legs, control box is not clicking.
•
Hand control is not properly plugged in.
•
Hand control is defective.
•
Control box is defective.
Hoist cannot spread or close legs, control box is clicking.
•
Leg spreading actuator cord (2) is not fully plugged into
slot (2) on the control box.
•
Leg spreading actuator is defective.
Cannot move hoist.
•
Rear caster brake is engaged.
•
Casters are defective.
Cannot tighten Mast into Base tube.
•
Mast locking assembly are defective, replace with new
Mast locking assembly .