background image

© Prior Scientific Instruments Ltd., UK 

Version 2.1 Nov 2014 

Page 

 

 

 

 

Make sure stage is in loading position, and a slide is correctly mounted onto the 
stage. Position the Oil Dispensing assembly close to the stage. IMPORTANT – MAKE 
SURE IT DOES NOT OBSTRUCT STAGE MOVEMENT 

 

Fasten down the Oil Dispenser assembly using the 3  screws provided. See below. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

Screw holes

 

 

 

 
 
 

 
 
 

 
 
 

 
 
 

The Dispensing tip should be positioned over the slide. To adjust position loosen the 
appropriate locking screws, and move tip into desired position. See for directions of 
movement. 

 

 
 
 

 
 
 

 
 
 

 
 
 

 
 

Adjustment 
screws

 

Summary of Contents for PL220

Page 1: ...Worldwide distribution Prior Scientific Ltd Prior Scientific Inc Prior Scientific GmbH Prior Scientific KK Cambridge UK Rockland MA USA Jena Germany Tokyo Japan T 44 0 1223 881711 T 1 781 878 8442 T...

Page 2: ...ENTIFYING THE PARTS OF THE PL220 CONTROLLER 7 3 2 IDENTIFYING THE PARTS OF THE DISPENSING TIP ASSEMBL 8 3 3 SET UP OF PL220 9 3 4 POSITIONING OF OIL DISPESNSER 10 3 4 INSTALLATION OF COM PORT 12 SECTI...

Page 3: ...able precise and valuable addition to your microscopy system Please do take the time to read these instructions before attempting to install and use this product This manual contains both important sa...

Page 4: ...r alter the AC cord or plug The power cord set must be an appropriately rated and approved cord set in accordance in the regulations of the country it is used in If the supplied plug adapter is not th...

Page 5: ...s can exceed normal operating range DO NOT run pump against a dead end condition closed discharge This can lead to excessive roller and tubing temperatures and pressures Rating The pump is rated as Gr...

Page 6: ...y return it to sales outlet in its original packaging No responsibility for damage arising from the use of non approved packaging will be accepted Ensure all items and accessories specified are presen...

Page 7: ...ific Instruments Ltd UK Version 2 2 November 2015 Page 7 INSTALLATION SECTION 3 3 1 Identifying the parts of PL220 Controller Front Direction of pump travel Pump Prime button RS232 Power connector USB...

Page 8: ...Version 2 2 November 2015 Page 8 3 2 Identifying the parts of Dispensing Tip Assembly Needle support arm Top clamp Dispensingtip Vertical support pillar Luer fitting Ankle adjustment locking screw Di...

Page 9: ...e the open end of the tubing into the bottle and ensure it is situated close to the bottom of the bottle and immersed in the oil Insert Dispensing Tip into holder and tighten locking screw Inset other...

Page 10: ...tely full of oil before releasing the PRIME button Unclamp the tubing from the pump and check that there are no air bubbles within the tubing inside the pump If bubbles are present continue to PRIME t...

Page 11: ...sing assembly close to the stage IMPORTANT MAKE SURE IT DOES NOT OBSTRUCT STAGE MOVEMENT Fasten down the Oil Dispenser assembly using the 3 screws provided See below Screw holes The Dispensing tip sho...

Page 12: ...il Dispenser is now ready to use 3 5 Installation of COM Port The RSR232 port is ready to use with standard PRIOR settings To install the USB port download the appropriate driver from the software dow...

Page 13: ...software Users should always read lines in until the END is seen A typical response isshown below OPTISCAN INFORMATION DRIVE CHIPS 00000 JOYSTICK IS DIGIPOT ONLY STAGE NONE FOCUS NONEL FILTER_1 NONE F...

Page 14: ...number of microsteps forward b number of microsteps backwards c number of microsteps per second when pumping with the PRIME button pushed DROP a 0 Delivers aspecified a number of DROPs Each DROP enta...

Page 15: ...Instruments Ltd UK Version 2 1 Nov 2014 Page REPLACEMENT PARTS SECTION 5 Part No Description H407 24v Desk top power supply W4013 Flexible Dispensing Tip W4014 Luer Fitting Screw W4040 Tubing 1m x 1...

Page 16: ...interference suppressors Therefore it is recommended that only a D C test be performed The remainder of the kit contains only low voltage DC circuits and should not be flash tested Ensure that the mai...

Page 17: ...is for indoor use and is designed to be safe within an ambient temperature range 5 to 40 C with maximum relative humidity of 80 relative humidity derated linearly to 50 rh above 31 C Dry weight 6 1Kg...

Page 18: ...cientific Instruments Limited declares that the following product PL220 Automated Oiler Conforms to the following Directives and Standards Machinery Directive 2006 42 EC EN809 1998 A1 2009 EN 61010 1...

Page 19: ...Please feel free to contact us regarding any questions comments or suggestions Prior Scientific manufactures a wide range of products designed for a huge range of microscopy applications from automat...

Reviews: