Chapter 5
Operation
59
59
4.
Verify that the slit is set to 10 µm.
If necessary, adjust the Exposure Time to
maintain optimum (near full-scale) signal intensity.
5.
Wait until the camera temperature locks at its default temperature.
6.
Make sure that the spectroscopy-mount adapter moves freely at the spectrograph
(you may need to loosen the two set screws that lock the sliding tube in place).
7.
Select
Align Spectrometer…
from the
Experiment Options
menu. Review the
displayed information and then click on the
Begin
button. Typically, this feature
creates three 1-row high ROIs (one near the top of the array, one in the middle,
and one near the bottom) and begins data acquisition. Data will be continuously
acquired and displayed but will not be stored.
Figure 30. Spectrometer Alignment: Before Rotational Alignment
8.
Adjust the rotational alignment. You do this by rotating the camera while
watching a live display of the line. Click on the peak you want to monitor during
the rotational alignment. This positions the large cursor to provide a vertical
reference line across all of the ROIs. Rotate the camera until the selected peak is
aligned horizontally in all of the ROIs.
Alternatively, take an image, display the horizontal and vertical cursor bars, and
compare the vertical bar to the line shape on the screen. Rotate the camera until
the line shape on the screen is parallel with the vertical bar.
Summary of Contents for ProEM+ EMCCD
Page 1: ...EMCCD Cameras For Imaging and Spectroscopy ProEM System Manual...
Page 2: ...4411 0144 Version 1 B January 07 2012 4411 0144...
Page 11: ...x ProEM System Manual Version 1 B This page intentionally left blank...
Page 17: ...16 ProEM System Manual Version 1 B This page intentionally left blank...
Page 77: ...76 ProEM System Manual Version 1 B This page intentionally left blank...
Page 85: ...84 ProEM System Manual Version 1 B Figure 46 Flowcharts of Safe and Fast Mode Operation...
Page 99: ...98 ProEM System Manual Version 1 B Figure 63 WinX Vertical Shift...
Page 121: ...120 ProEM System Manual Version 1 B This page intentionally left blank...
Page 125: ...124 ProEM System Manual Version 1 B This page intentionally left blank...
Page 127: ...126 ProEM System Manual Version 1 B ProEM 512BK Kinetics Figure 84 ProEM 512BK Outline Drawing...
Page 130: ...Appendix B Outline Drawings 129 129 CoolCUBEII optional Figure 87 CoolCUBEII Outline Drawing...
Page 131: ...130 ProEM System Manual Version 1 B This page intentionally left blank...