20
ProEM
®
+ System Manual
Version 1.B
Coolant Circulator
ProEM+ cameras can be optionally cooled by circulating coolant to provide a low
vibration system for data acquisition. They are equipped with two ports with quick-
disconnect fittings for connection to an external circulator. Two 10 mm (3/8”) ID, 3
meter (10 ft) coolant hoses with ProEM+-compatible fittings are provided with each
ProEM+ system.
The coolant circulator can be any commercially available circulator (provided that it is
capable of continuously pumping a 50:50 mixture of room temperature (23ºC) water and
ethylene glycol at 1 liter per minute) or the Princeton Instruments CoolCUBE
II
. Please
contact the factory for recommendations.
Cables
Ethernet Cable:
This standard 5 meter (16.4’) cable is a Cat 5e/6 Ethernet
cable (6050-0621) for interconnecting the camera and the host computer. The
distance between the camera and the computer can be over 50 meters. Please
contact the factory for longer cables.
MCX to BNC Adapter Cables:
Two MCX to BNC adapter cables
(6050-0540) are provided with the ProEM+ system. These connect to the EXT
SYNC and the LOGIC OUT connectors on the rear of the ProEM+.
Power Cable:
The power cable (6050-0596) interconnects the ProEM+
camera and its power supply. This cable is approximately is 3 m long
(approximately 10 ft).
Hoses
Coolant Hoses:
Two 10 mm (3/8”) ID, 3 meter (10’) hoses are supplied with
every ProEM+ system. Quick-disconnects that mate to the ProEM+’s coolant
ports have been installed on one end of each hose. Refer to your coolant
circulator’s specifications regarding circulator-compatible hose fittings. If a
Princeton Instruments CoolCUBE
II
circulator is ordered with the camera,
hoses are supplied with appropriate connectors on both ends.
Note:
Part numbers for the hose, ProEM+ fitting, and CoolCUBE
II
fitting are
McMaster # MCM 5238K748, McMaster # MCM 5478K811, and CPC #
NS6D17006, respectively.
Certificate of Performance
Each ProEM+ camera has a Certificate of Performance. This certificate states that the
camera system was assembled and tested according to approved Princeton Instruments
procedures. It documents the camera performance data as measured during the testing of
your ProEM+ and lists the Sales Order, Purchase Order, and Camera Serial numbers
(useful if you ever need to contact Princeton Instruments Customer Support).
Summary of Contents for ProEM+ EMCCD
Page 1: ...EMCCD Cameras For Imaging and Spectroscopy ProEM System Manual...
Page 2: ...4411 0144 Version 1 B January 07 2012 4411 0144...
Page 11: ...x ProEM System Manual Version 1 B This page intentionally left blank...
Page 17: ...16 ProEM System Manual Version 1 B This page intentionally left blank...
Page 77: ...76 ProEM System Manual Version 1 B This page intentionally left blank...
Page 85: ...84 ProEM System Manual Version 1 B Figure 46 Flowcharts of Safe and Fast Mode Operation...
Page 99: ...98 ProEM System Manual Version 1 B Figure 63 WinX Vertical Shift...
Page 121: ...120 ProEM System Manual Version 1 B This page intentionally left blank...
Page 125: ...124 ProEM System Manual Version 1 B This page intentionally left blank...
Page 127: ...126 ProEM System Manual Version 1 B ProEM 512BK Kinetics Figure 84 ProEM 512BK Outline Drawing...
Page 130: ...Appendix B Outline Drawings 129 129 CoolCUBEII optional Figure 87 CoolCUBEII Outline Drawing...
Page 131: ...130 ProEM System Manual Version 1 B This page intentionally left blank...