DAILY USAGE
Monitor indicators
*
The monitor lights are faint. They can best be detected in a dark or dimly lit room.
Sounds
Lights*
What it means
Caused by
none
every 5 sec Monitoring successfully.
and
none
Symptom marked.
Press-and-release button push.
and
every 1 sec Poor skin contact.
Monitor cannot detect good contact with skin to begin monitoring (
p. 11, 16
).
and
every 5 sec Poor skin contact.
Monitor has not detected good contact with skin for over 8 hours.
Replace electrodes.
none
and
every 5 sec Poor skin contact.
Monitor cannot detect good contact with skin during already established
monitoring session. If this persists for 30 min or more, replace
electrodes.
and
every 5 sec Monitor battery low.
Low monitor battery. Charge the monitor (
p. 18
).
and
Monitor powering on.
Press-and-release button push.
and
Monitor powering off.
Press-and-hold (12-sec) button push.
After 12 seconds, all four monitor lights flash once, followed by a long beep.