16
GETTING STARTED
for patients using an
ECG LEAD SET
Turn on monitor
Press and release the center button on the monitor once.
If contact between the ECG lead set and your skin is satisfactory:
The monitor makes a quick series of three rising beeps and the green light blinks every 5 seconds.
You are
successfully monitoring.
If contact between the ECG lead set and your skin is NOT satisfactory
:
The monitor alerts you with two beeps every second. If this occurs:
1. Press and hold the center button on the monitor for 12 seconds. After 12 seconds, the monitor lights will
flash once, followed by a long beep. This indicates that the monitor has powered off.
2. Revisit the previous steps in this manual to ensure that:
• The monitor and leads are securely attached to the ECG electrodes (
page 13
).
• BodyGuardian Mini is properly placed and securely adhered to your chest (
pages 14-15
).
3. Power on the monitor by pressing and releasing the center button on the monitor once.
• If the monitor does not beep continuously, you are successfully monitoring.
• If the monitor beeps twice every second, remove the electrodes, use new electrodes, and begin again on
page 12.