PRESYS
Instruments
PCON-Y17
Page 13
WARNING!
The calibrator discharges gas to the atmosphere through the supply (-)
and vent ports. Leave these ports free.
The calibrator may produce high sound levels when discharging gas. Use
a muffling device at the vent port.
WARNING!
Before connecting the instrument under test to the output port, make a
VENT operation to discharge any gas inside the controller and go to
MEASURE mode. Before turning the calibrator off, it is recommended to
vent the gas inside the controller.
IMPORTANT!
All pressure equipment and accessories (such as hoses, connections,
adapters, etc.) connected to the calibrator must be clean, free of residues
such as oil, dirt, dust etc. These residues can cause irreversible damage
to the internal system of the pressure control module. It is mandatory to
use filters in both the pressure feed (SUPPLY
(+)
/ SUPPLY
(-)
) and out
(OUT).
presys