Page 51
2.3.2. Performing Tasks
To perform an existing task, select
TASKS
>
EXPLORE TASKS
from the main
menu. A list identifying the created work orders that have not been performed yet
(●
WAITING
) is shown. Select the desired task and press
OK
. Make the necessary
connections between the calibrator and the instrument to be calibrated and press
START
.
Fig. 69
- Exploring tasks
PCON-Y17 automatically starts the calibration by generating the set points
registered in the task and doing the reading of the instrument to be calibrated. If you
select the
NONE
option for the input, for each generated point the calibrator requires the
value read by the instrument.
For inverted calibrations, it is possible to vary the control set point in small steps,
until the pointer of the analog gauge indicates the desired value. When the initial
set point stabilizes, the following screen is shown. This set point variation is made
through the buttons “+” and “-” and the steps shall be established by the option
STEP.
When the analog gauge reaches the desired value, press
OK
to confirm the point.
Fig. 70
- Inverted Calibration