PD6800 Loop-Powered Process Meter Instruction Manual
18
Main Menu Display Functions & Messages
The meter displays various functions and messages during setup, programming, and
operation. The following table shows the main menu functions and messages in the
order they appear in the menu.
Display Parameter
Action/Setting
SETUP
Setup
Enter
Setup
menu
DeC..pt
Decimal point
Enter
Decimal Point
menu
PRog
Program
Enter the
Program
menu
sCalE
Scale
Enter the
Scale
menu
Cal
Calibrate
Enter the
Calibrate
menu
Inpt1
Input 1
Calibrate input 1 signal or program input 1 value
DspL1
Display 1
Program display 1 value
Inpt2
Input 2
Calibrate input 2 signal or program input 2 value
DsPl2
Display 2
Program display 2 value
Span
Error
Span
Error
Error, calibration not successful, check signal
tag
Tag/Units
Enter the Tag/Units Menu
ON
Tag On
Enable Tag/Units
OFF
Tag Off
Disable Tag/Units
PASSWRD
Password
Enter the
Password
menu
UNLOCKD
Unlocked
Program password to lock meter
LOCKED
Locked
Enter password to unlock meter
99999
-99999
Flashing
display
Overrange condition
Underrange condition
SERVICE
Service
Select before removing/installing cover for ser-
vice or to clean the glass window