– 10 –
To use RECAL option:
Step 1:
Press the Menu key on the keypad
Step 2:
Press the
(up arrow key) or
(down arrow
key) on the keypad until the word CONFIG appears.
Step 3:
Press Enter.
Step 4:
Enter user password if it has been programmed.
Step 5:
Press
(up arrow key) or
(down arrow key)
until the word RECAL appears.
Step 6:
Press Enter.
Step 7:
The gauge will now flash between INPUT and unit
of measure on the lower line and .00 on the top line. Apply
zero pressure to the gauge.
Step 8:
Press Enter. Zero pressure is now set.
Step 9:
The gauge will display full-scale pressure. Apply
full-scale pressure to the gauge.
Step 10:
Press Enter. Full-scale pressure is now set.
Step 11:
The gauge will now display mid-scale pressure.
Apply mid-scale pressure to the gauge.
Step 12:
Press Enter. Mid-scale pressure is now set.
(Note: For compound ranges this recalibration is zero,
full-scale, mid-scale and full-vac.)
FOR FACTORY CALIBRATED SETINGS:
Step 13
: To reinstate factory calibrated settings for zero,
full-scale and mid-scale press the
(down arrow key)
Menu key until the word FACT appears.
Step 14:
Press Enter. Factory calibration settings are now
reinstated.
Step 15:
After zero, full-scale and/or mid-scale or factory
default calibration have been set, the word SAVE appears
on the gauge display.
Step 16:
Press Enter to finalize calibration.
MENU FUNCTIONS
*Indicates default.
I&M008-10136-DIG PDC 7-05_layput 8/4/11 2:42 PM Page 10