PD662 Loop-Powered Meter
Instruction Manual
15
Setting Numeric Values
The numeric values are set using the
Right
and
Up
arrow buttons.
Press
Right
arrow to select next digit and
Up
arrow to increment digit.
The two left-most digits on the display are set as a single digit, able to
display -19 to 29.
The digit being changed blinks.
Press the
Enter
button, at any time, to accept a setting or
Menu
button
to exit without saving changes.
The decimal point is set using the
Right
or
Up
arrow button in the
Setup-decimal point
menu.
Programming the Meter
It is
very important
to read the following information, before proceeding
to program the meter:
There is
no need to recalibrate
the meter for milliamps when first
received from the factory.
The meter is
factory calibrated
for milliamps prior to shipment.
The calibration equipment is certified to NIST standards.
Use the
Scale
(
SCL
) menu to enter scale parameters without
applying a live signal.
Alternatively, use the
Calibrate
(
CAL
) menu to apply a signal from a
calibrator or a 4-20 mA transmitter to calibrate the meter.
Inputs may be calibrated or scaled to any display within the range of the
meter.
Additional parameters, not needed for most applications, are viewed
and programmed with the
Advanced
Features
menu, see Advanced
Features Menu on page 20.