PD662 Loop-Powered Meter 

Instruction Manual 

 

 

 

 

 

4-20 mA Input Loop-Powered 

 

-1999 to 2999 Display 

 

Easy Four-Button Programming 

 

NEMA 4X Enclosure 

 

Loop-Powered Backlight Option 

 

1.7 Volt Drop without Backlight 

 

32-Point and Square Root Linearization Input Functions 

 

Programmable Noise Filter 

 

HART Protocol Transparent 

 

CSA Certified for -40 to 75°C Operation 
(-55 to 75°C with Heater Option) 

 

 

C             US

Summary of Contents for Survivor PD662

Page 1: ...uction Manual 4 20 mA Input Loop Powered 1999 to 2999 Display Easy Four Button Programming NEMA 4X Enclosure Loop Powered Backlight Option 1 7 Volt Drop without Backlight 32 Point and Square Root Linearization Input Functions Programmable Noise Filter HART Protocol Transparent CSA Certified for 40 to 75 C Operation 55 to 75 C with Heater Option C US ...

Page 2: ...re malfunctioning could result in personal injury or property loss Anyone using this product for such applications does so at his her own risk Precision Digital Corporation shall not be held liable for damages resulting from such improper use Limited Warranty Precision Digital Corporation warrants this product against defects in material or workmanship for the specified period under Specifications...

Page 3: ...i point linearization with up to 32 points and the square root function allow for conditioning of signals from non linear transmitters without adding external components to the system The PD662 is housed in a rugged NEMA 4X enclosure and is provided with one NPT pipe conduit hole The PD662 is also available with an optional DC powered heater that allows the PD662 to operate in temperatures as low ...

Page 4: ...alues 15 Programming the Meter 15 Main Menu 16 Main Menu Display Functions Messages 17 Setting the Decimal Point dP 17 Scaling the Meter SCL 18 Calibrating the Meter Cal 19 Recalibrating the Internal Calibration Reference Ical 19 Advanced Features Menu 20 Advanced Features Menu Display Messages 21 Signal Input Conditioning Function Fnc 22 Information Menu info 22 Input Signal Filter FIL 22 Interna...

Page 5: ...NFORMITY 33 Table of Figures Figure 1 PD662 Mounting Board Rear View 11 Figure 2 PD662 Input Connections without Backlight 12 Figure 3 PD662 Input Connections with Backlight 12 Figure 4 PD662 Heater Power Connections 12 Figure 5 Case Dimensions Front View 28 Figure 6 Case Dimensions Side View 28 Figure 7 Case Dimensions Bottom View 29 Figure 8 PDA6845 Pipe Mounting Kit 29 ...

Page 6: ... a minimum of ten years if power is lost NORMAL MODE REJECTION 64 dB at 50 60 Hz ENVIRONMENTAL Operating temperature for CSA certification 40 to 75 C Functional temperature range 40 to 85 C Storage temperature range 40 to 85 C Relative humidity 0 to 90 non condensing CONNECTIONS Removable screw terminals accept 12 to 22 AWG wire ENCLOSURE Impact resistant polyester plastic body gray cover clear wi...

Page 7: ...ANCE 85 Ω 20 mA 185 Ω 20 mA INPUT OVERLOAD Over current protection to 2 A max Heater POWER 24 VDC 11 W max REQUIRED SWITCH FUSE A CSA Approved fuse rated min 30 VDC 1 A T and a readily accessible switch or a circuit breaker shall be provided to facilitate the disconnection of power to the heater OPERATION METHOD On Off control with an internal temperature sensor Heater is on only when necessary du...

Page 8: ... Laboratory Use Part 1 General Requirements UL 61010 1 2004 2nd Edition Safety Requirements for Electrical Equipment for Measurement Control and La boratory Use Part 1 General Requirements LOW VOLTAGE DIRECTIVE EN 61010 1 2010 Safety requirements for measurement control and la boratory use IEC 61010 1 2010 Safety requirements for measurement control and la boratory use Electromagnetic Compatibilit...

Page 9: ...sonnel This unit is not suitable for hazardous locations such as potentially explosive zones or locations No user replaceable part inside All service shall be performed at the factory Do not disconnect and reconnect cables during a lightning storm Suitable surge suppressor e g lightning protection shall be provided in end application installation INSTALLATION All pushbuttons and wiring connectors ...

Page 10: ...e conduit and a CSA Approved connector rated 4X is intended to be used in final installation Metallic conduit shall be reliably bonded to protective earth ground without creating any ground loop PD662 Connections Disconnect power before opening the unit To access the connectors remove the enclosure cover and unscrew the two captive stainless steel screws on the rear mounting board Remove the meter...

Page 11: ...l S Connect signal negative to terminal S If backlight installed connect signal negative to t erminal B not S S S B S S B Figure 1 PD662 Mounting Board Rear View Connection Wiring Diagrams Signal input connections are made to a three terminal connector labeled S S X for models without a backlight and S S B for models with a backlight The 4 20 mA input with no backlight has a maximum voltage drop o...

Page 12: ...ons with Backlight Figure 4 PD662 Heater Power Connections A CSA Approved fuse rated min 30 VDC 1 A T and a readily accessible switch or a circuit breaker shall be provided to facilitate the disconnection of power to the heater 1 2 24 VDC 11 W max P P S S X 4 20 mA Transmitter Power Supply S S B 4 20 mA Transmitter Power Supply ...

Page 13: ...actory calibrated for milliamps prior to shipment The calibration equipment is certified to NIST standards Overview There are no jumpers involved in the setup process of the meter Setup and programming is done through the front panel buttons After all connections have been completed and verified apply power to the loop For a Quick User Interface Reference Guide go to page 31 ...

Page 14: ...s of the meter ENTER Enter button to access a menu or accept a setting Right arrow to scroll through the menus or move to the next digit or decimal position during programming Displays the Max then Min display values when pressed during normal run mode Up arrow to scroll through the menus decimal point or to increment the value of a digit Resets the Max and Min display value when pressed during no...

Page 15: ...It is very important to read the following information before proceeding to program the meter There is no need to recalibrate the meter for milliamps when first received from the factory The meter is factory calibrated for milliamps prior to shipment The calibration equipment is certified to NIST standards Use the Scale SCL menu to enter scale parameters without applying a live signal Alternativel...

Page 16: ...L Run Mode MENU Press Menu button to enter Programming Mode then press the Up arrow button to scroll main menu Press Menu at any time to exit and return to Run Mode Changes made to settings prior to pressing Enter are not saved Changes to the settings are saved to memory only after pressing Enter The display moves to the next menu every time a setting is accepted by pressing Enter ...

Page 17: ...ed value for input 1 in2 Scale Input 2 Input signal 2 value mA D2 Scale Display 2 Scaled value for input 2 CAL Calibrate Enter the Calibrate menu Npt Number of Points Set number of linearization points in1 Calibrate Input 1 Read input signal 1 D1 Calibrate Display 1 Enter value for input 1 in2 Calibrate Input 2 Read input signal 2 D2 Calibrate Display 2 Enter value for input 2 Setting the Decimal ...

Page 18: ...sed in the Scale menu 2 to 32 points may be used The Scale menu is entered after entering the num ber of points Minimum Input Span The minimum input span is the minimum difference between input 1 and input 2 signals required to complete the calibration or scaling of the meter The minimum span is 0 40 mA If the minimum span is not maintained the meter reverts to input 2 allowing the appropriate inp...

Page 19: ...al input and press Enter to accept 5 The meter displays in2 Apply a known signal and press Enter The display will flash while accepting the signal 6 When the meter displays d2 press Enter Enter a corresponding display value for the signal input and press Enter to accept Recalibrating the Internal Calibration Reference Ical The Internal Calibration ICAL menu located in the Advanced Features menu is...

Page 20: ...ns are located in the Advanced Features menu Press and hold the MENU button for five seconds to access the Advanced Features menu Press Enter to Access Menu or to Accept Setting ICAL FIL ENTER MENU Press Up to Scroll Menu and to Increment Digit Value Press Menu to Exit at any Time inFo MENU Press and hold for five seconds 26 62 Run Mode fnc ...

Page 21: ...unction Lnr Linear Set linear scaling Sqr Square Root Set square root input conditioning func tion info Information Enter the Information menu Sft Software Information Software release number ver Version Meter firmware version C Calibration Temp C Temperature at time of I calibration C F Calibration Temp F Temperature at time of I calibration F FIL Filter Set filter function level ICAL I Calibrati...

Page 22: ...emperatures To determine the software version of a meter 1 Go to the Information menu info and press Enter button 2 The meter will automatically scroll through the software release number and software version The meter temperatures at the time of last internal calibration in C and F are displayed for calibration troubleshooting Pressing the ENTER RIGHT or UP buttons will progress the information d...

Page 23: ...bration procedure The Internal calibration menu is part of the Advanced Features menu 1 Press and hold the MENU button for 5 seconds to enter the Advanced Features menu 2 Press the Up arrow button to scroll to the Internal Calibration menu ICAL and press Enter 3 The meter displays 4 00 mA Apply a 4 00 mA signal and press Enter The display flashes for a moment while the meter is accepting the signa...

Page 24: ...ttons Operation Button Symbol Description MENU Press to enter or exit Programming Mode or exit Max Min readings Press to reset Max Min readings Press to display Max Min readings alternately ENTER Press to display Max or Min reading indefinitely while displaying Max or Min ...

Page 25: ...in 19 45 5 34 1 Press Up arrow button to display maximum reading since the last reset power up 2 Press Up arrow again to display the minimum reading since the last reset power up 3 Press Enter to continue to display the Max or Min display reading by disabling the Max Min timeout The meter will continue to track new Max Min readings Press MENU to exit the Max Min reading 4 If Enter is not pressed t...

Page 26: ...utton when rSt is shown Press Enter again when display flashes rSt Note If Enter is not pressed a second time within three seconds rSt will stop flashing and the last ENTER press cancelled 5 The meter goes through an initialization sequence same as on power up and loads the factory default settings Heater Operation Models with the integrated heater option require no additional setup All heater ope...

Page 27: ...tory setting the user may record the new setting for the particular application Model ______________ S N _______________ Date _________ Parameter Display Default Setting User Setting Decimal point Dd dd 2 places Scale SCL Number of Points NPT 2 Input 1 in1 4 00 mA Display 1 D1 4 00 Input 2 in2 20 00 mA Display 2 D2 20 00 Advanced Features Input Conditioning Function Fnc Linear Filter FIL Off ...

Page 28: ...UNTING DIMENSIONS Figure 5 Case Dimensions Front View Figure 6 Case Dimensions Side View A 3 15 in 80 mm B 4 33 in 110 mm C 2 44 in 62 mm D 3 62 in 92 mm E 2 76 in 70 mm 3 35 in 85 mm with heater option F 0 79 in 20 mm 1 38 in 35 mm with heater option ...

Page 29: ...ttom View G 0 87 in 22 mm H 0 80 in 20 mm I 2 17 in 55 mm Figure 8 PDA6845 Pipe Mounting Kit The PDA6845 pipe mounting kit may be used to mount a PD662 to a 2 pipe The kit includes an 8 20 in 208 mm x 4 10 in 104 mm mounting panel with a 2 U bolt and all necessary mounting hardware ...

Page 30: ...ter displays error message during calibration err Check 1 Signal connections 2 Minimum input span requirements Meter flashes 2999 1999 Check 1 Input signal within scaled range of 2999 and 1999 Display stuck flashing a number and HI or LO Press Menu to exit Max Min display readings Display response is too slow Check filter setting to see if it can be lowered to LO or OFF If the display locks up or ...

Page 31: ...Accept selection value and move to next selection Max Min Mode While in Run Mode pressing Up Arrow will initiate Max Min Mode Up Arrow toggles between Max Min displays and Right Arrow resets the Max Min to the current value Press Menu or wait 10 seconds to return to Run Mode Pressing Enter will disable the 10 second timeout and continuously flash Max or Min Decimal d2 Program Run Max Min Advanced ...

Page 32: ...PD662 Loop Powered Meter Instruction Manual 32 This Page Intentionally Left Blank ...

Page 33: ...nce with the application of harmonized or applicable technical standards and when applicable or required a European Union notified body certifi cation Standards EN 61010 1 2001 EN 61326 2006 The standards EN 61010 1 2001 and EN 61326 2006 are no longer harmonized The requirements of these standards have been checked against the harmonized standards EN 61010 1 2010 and EN 61326 2013 and there were ...

Page 34: ...p 08 18 How to Contact Precision Digital For Technical Support Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com For Sales Support Call 800 343 1001 or 508 655 7300 Fax 508 655 8990 For the latest version of this manual please visit www predig com ...