PD6928 Ex-Proof & I.S. Loop-Powered Flow Rate/Totalizer
Instruction Manual
7
Key Features
2X More Informative Display
The PD6928 display offers a 50% larger display area
and is twice more informative than previous
generations of loop-powered meters. Featuring an
alphanumeric dual-line display and a 20-segment
bargraph, reading and understanding process values
is easy and intuitive. The addition of status indicators
provides a quick glance at alarm conditions, relays,
process trends, and more. Predefined display units
give users even more display flexibility. Plus, the high
contrast backlit LCD display is readable from far away
and under various lighting conditions.
Backlight Turns Red on Alarm
When an alarm occurs, the display can be
programmed to turn red, flash, and display an alarm
indicator
. In addition, a unique custom alarm
message for each of the two relays and two open
collectors can be displayed on the bottom display.
These features can be activated even if no relay or
open collector is connected.
CapTouch Through-Glass
Buttons
The PD6928 is equipped with four capacitive sensors
that operate as through-glass buttons so that it can be
operated without removing the cover (and exposing
the electronics) in a hazardous area. CapTouch
buttons are designed to work under any lighting
condition and to protect against false triggering. They
can be disabled for security via a switch on the
display module.
CapTouch buttons have two modes of operation:
Normal and Delayed. Delayed mode prevents
accidental trigger of the buttons. In the Delayed
mode, the buttons enter into a low sensitivity state
(sleep) and they ignore quick button presses after 20
seconds of inactivity. To wake up the buttons, press
and hold any button for more than 2 seconds, the
buttons respond normally.
Solid State Relays 1-2
Indicators
Predefined
Rate Units
0.7" (17.0 mm) High
5-Character, Alphanumeric
12-segment
Password Protection
Trend
Indicators
CapTouch Buttons
Self-Calibration
0.4" (10.2 mm) High
8-Character, Alphanumeric
14-segment
Total or Grand Total
Indicator
Alarm Indicator
20-Segment
Bargraph
Two-Color
Backlight
CapTouch Buttons /
Function Keys