Specifications
Except where noted all specifications apply to operation at +25°C.
Display
Display
Dual-line LCD with backlight.
Top: 2.8" (71 mm),
5 alphanumeric 14-segment characters.
Bottom: 1.5" (39 mm),
8 alphanumeric 14-segment characters.
Display may be programmed to turn red
and flash a user-defined message on
alarm condition.
Top Display
5 digits (-9999 to 99999) or 5 characters
(all capital & most lower-case letters)
Bottom Display
8 digits (-9,999,999 to 99,999,999;
separated by commas) or 8 characters
(all capital & most lower-case letters)
Backlight Power
Requirement
24 VDC @ 46 mA, typical
Bargraph
20 segments, numeric percent indication
at top
Decimal Point
Up to four decimal places on top display and
up to seven decimal places on bottom display
Commas
Commas to indicate 1000s
(e.g. 88,987,628) on bottom display only
Dual-Scale
Feature
The input can be displayed in different
scales on the top and bottom displays.
For instance, the top display could display
the flow in GPM and the bottom display
could display that same input in CFM.
Alarm Indication
Programmable: red backlight, flashing
display, Bargraph segment flashes on alarm.
Backlight requires external 24 VDC.
Custom Alarm
Messages
Programmable for each relay/open collector:
8 characters maximum; displayed every 10 sec
for 1 sec on bottom display. May be turned off.
Display Update
Rate
Ambient > -10°C: 1 Update/Second
Ambient = -20°C: 1 Update/2 Seconds
From -20°C to -40°C the update rate
slows down 1 second for every -2°C
(e.g. at -24°C, 1 update/4 seconds).
Overrange
Top: 99999; Bottom: 99,999,999 (flashing)
Underrange
Top: -9999; Bottom: -9,999,999 (flashing)
General
Programming
Method
Buttons behind lower panel door & Free
PC-based USB programming software
Enclosure &
Materials
Material: High impact Polycarbonate with
UV stabilizer enclosure, UL 94V-0
Rating: NEMA 4X / IP65
Gasket: Polyurethane
Color: gray
Includes four PG11 through-hole conduit
openings, with two factory installed PG11,
IP68, black nylon threaded hole plugs with
backing nuts.
Environmental
Operating temperature range:
-40 to 75°C (-40 to 167°F)
Storage temperature range:
-40 to 85°C (-40 to 185°F)
Relative humidity: 0 to 90% non-condensing;
Printed circuit boards are conformally coated.
Noise Filter
Averages the input signal over a period of
time between 1 and 16 seconds to
dampen the effects of a noisy signal that
causes a jumpy display.
Filter Bypass
0.0 to 99.9% of full scale. Input signal
changes greater than bypass value are
displayed immediately.
Recalibration
Recalibration is recommended at least
every 12 months.
Max/Min
Display
Max/min readings reached by the process
are stored until reset by the user or until
power to the meter is turned off.
Tare
If the totalizer is disabled, the Tare
function zeros out the meter to remove
the weight of a container. Tare function
can be assigned to a function key or the
digital input.
Password
There are three separate passwords
available that can be set independently of
each other: Main, Total, and Grand Total.
The Main password prevents access to
the meter Programming Mode. Total and
Grand Total passwords prevent resetting
the total and grand total, respectively.
Non-Volatile
Memory
Total and Grand Total values, and all
programmed settings are stored in non-
volatile memory for a minimum of ten
years if power is lost.
Normal Mode
Rejection
64 dB at 50/60 Hz
Connections
Removable screw terminals accept 12 to
22 AWG wire
Remote contacts: Accept 16 to 30 AWG wire.
Tightening
Torque
Screw terminal connectors: 4.5 lb-in (0.5 Nm)
Mounting screws: 8.0 lb-in max. (0.9 Nm)
Remote contacts: 2.5 lb-in (0.28 Nm)
Overall
Dimensions
10.6" x 12.6" x 4.8"
(270 mm x 320 mm x 121 mm) (H x W x D)
Weight
5.5 lbs (2.5 kg)
Warranty
3 years parts and labor. See Warranty
Information and Terms & Conditions on