4
Safety certifications: CE directive 2006/95/EC and 2004/108/EC, standard used EN61010-1, IEC61010-1,
EN 61326-
1:2006,
IEC 61326-1:2005
Installation location conditions
Operation site: Indoors
Maximum operating altitude: 2500 m or lower
Operating temperature 3ºC to 42ºC
WARNING:
(1) Do not operate the instrument under voltage fluctuations exceeding 10% of the recommended line voltage. Large
fluctuations may cause the instrument to fail. Use a three-pronged electrical outlet with a ground.
(2) Instrument can be used in the temperature range 3°C - 42°C, avoid freezing. Do not install the equipment at a place
where the temperature changed frequently.
Instrument can be used under a humidity range of 30 - 80% (RH). Relative humidity less than 80% from 3ºC to 30ºC,
decreasing linearly to 50% from 31ºC to 42 ºC
(3) Do not install the equipment near a heating element.
(4) Do not install the equipment at a place where it may be exposed to corrosive gas.
(5) Do not install the instrument in a location where it may be exposed to dust, especially in locations exposed to outside
air or ventilation outlets that discharge dust particles.
(6) Do not install the equipment at a place constantly or excessively exposed to oscillations or impacts.
(7) Do not install the instrument in a location where it may be exposed to direct sunlight.
(8) Avoid strong magnetic fields and sources of high-frequency waves. The instrument may not function properly near
strong magnetic fields or high frequency wave sources.
The WEEE (Waste Electrical and Electronic Equipment) symbol indicates that this product should not be
disposed of in unsorted municipal waste. Follow local municipal waste ordinances for proper disposal
provisions to reduce the environmental impact of WEEE.