18
Replacing the Fuse
Follow the instructions below to replace the 250V, 5A rated fuse for the power socket.
1. Turn off BlotCycler using switch on the back of the instrument and detach the power cord from the rear of
the instrument.
2. Open the fuse compartment located on the power entry block using a small flat blade screwdriver or
fingernail to gently open the fuse compartment.
3. Pull the fuse holder out of the compartment and inspect the fuse. If the fuse is burned or there is a break in the
fuse element, replace the 250 V, 5 A with the identical type fuse.
4. Place the fuse holder back into the compartment and snap the cover closed.
For additional fuses, contact Customer Support.