5 Operation
■
43
5.5.2 The print key
– Ensure that there is no drying taking place, i.e. that the instrument
is in weighing mode.
•
Printing out an individual value or a report
– Press «
» briefly.
– The drying report is printed out. The report for the last
measurement can be printed out at the start of a new drying
procedure. Reports printed afterwards do not contain any interim
results. It is otherwise identical to the drying report which is
printed during the measurement. If no drying has been performed
since the instrument was started up, the weight value is printed.
•
Printing out the statistics (only XM 60, XM 60-HR)
– Keep «
» pressed until "PRINT STATISTICS" is displayed
– Release «
»
– Statisics are printed.
•
Resetting the counter to 1 (only XM 66)
– Keep «
» pressed until "RESET COUNTER" is displayed.
– Release «
».
– The counter is reset to 1.
•
Printing the instrument settings
– Keep «
» pressed until "PRINT STATUS" is displayed.
– Release «
». The instrument settings are printed.
Status :
--------
Status printout for the settings.
Date 07.10.2016 Time 16:12:39
Name
: XM 60
Software
: C01-0000 P06
Serienr
: 5600500
Instrument identification
Print :
Printformat :
Date and Time
: on
Balance-ID
: on
Method-ID
: on
Counter
: on
Dryer Setup
: on
Print rate
: off
Operator-ID
: off
Print rate
: 1.0 min
Operator
:
Mode
: Printer
Report printout settings
Calibration :
Mode
: external
Defined weight
: 0.0000 g
Balance calibration settings
Temp. cal.
: off
Temperature calibration settings