3 Set up
14
• Open the hood and insert the draftshield (1), making sure that it is
placed on flat.
• Insert the pan support (2) and turn it so that it locks securely in
place.
• Insert the sample holder (3) as shown.
• You can now place an aluminium pan (4) on the pan support.
3.4
Choosing a suitable location
In order to ensure that the moisture analyser functions properly, select
a location that fulfils the following criteria:
• Permissible ambient temperatures
– Temperature: 5°C - 40°C
– Relative humidity: 25% - 85%, non-condensing
• Put the instrument on a rigid, firm horizontal base, preferably
exposed to no vibrations
• Make sure that the instrument cannot be shaken or knocked over
• Do not expose it to direct sunlight and dusy environment
• Avoid drafts and excessive temperature fluctuations
• Leave enough clear space around the instrument to prevent a build-
up of heat.
Do not expose the instrument to high levels of moisture for long
periods of time. Avoid letting condensation form on the instrument. If
instruments are cold, let them warm up to room temperature (approx.
20°C) before connecting them to the mains.
Condensation is practically impossible on instruments which are
connected to the mains.
NOTE
All the parts must be attached together without exerting undue
force. Do not apply any force. If you have any problems, the Precisa
customer service will be happy to help.