
■
6 Determining moisture levels
54
6.3.4 Switch-off criteria
This function key is used to start the definition of the switch-off
criteria for the measurement.
XM 60, XM 60-HR
Two freely definable stop modes are available according to the "digit
per time" resp. "% per time" principle. There is also the automatic
stop mode "ADAPTSTOP".
XM 66
There are five fixed and one freely definable stop modes available
according to the "digit per time" principle. There is also the automatic
stop mode "ADAPTSTOP".
• STOPMODE
DIGIT/TIME
2/20
Only if Digit/Time is selected
%/TIME
0.2/20
Only if %/Time is selected
AUTO STOP
DIGIT/TIME
%/TIME
ADAPTSTOP
OFF
• STOPMODE
AUTOSTOP
01/20
D/S
only if FREE is selected
AUTOSTOP
OFF
2/10
2/20
2/30
2/60
10/60
ADAPTSTOP
FREE